  "version": "2026-11-20",
  "publisher": "Vora",
  "method": "Manual extraction from Frija-Masson et al. main Results, preserving signed values, IQR, model identity and cross-sectional limits. Separate original synthetic example.",
  "rows": [
    {
      "record_type": "published_result",
      "entity": "Tefal Body Partner",
      "n": 53,
      "metric": "weight",
      "median": 0.3,
      "q1": -0.1,
      "q3": 0.7,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "published_result",
      "entity": "Tefal Body Partner",
      "n": 53,
      "metric": "fat mass",
      "median": -2.2,
      "q1": -5.8,
      "q3": 1.3,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "published_result",
      "entity": "Tefal Body Partner",
      "n": 53,
      "metric": "muscular mass",
      "median": 4.5,
      "q1": 0.4,
      "q3": 7.3,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "published_result",
      "entity": "Terraillon DietPack",
      "n": 52,
      "metric": "weight",
      "median": 0,
      "q1": -0.4,
      "q3": 0.3,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "published_result",
      "entity": "Terraillon DietPack",
      "n": 52,
      "metric": "fat mass",
      "median": -4.4,
      "q1": -6.6,
      "q3": 0,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "published_result",
      "entity": "Terraillon DietPack",
      "n": 52,
      "metric": "muscular mass",
      "median": -6.6,
      "q1": -9.4,
      "q3": -3.6,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "published_result",
      "entity": "Nokia Withings Body Cardio",
      "n": 48,
      "metric": "weight",
      "median": 0.25,
      "q1": -0.1,
      "q3": 0.5,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "published_result",
      "entity": "Nokia Withings Body Cardio",
      "n": 48,
      "metric": "fat mass",
      "median": -3.7,
      "q1": -8,
      "q3": 0.3,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "published_result",
      "entity": "Nokia Withings Body Cardio",
      "n": 48,
      "metric": "muscular mass",
      "median": 4,
      "q1": 0.1,
      "q3": 7.6,
      "unit": "kg",
      "region": "French tertiary hospital",
      "source_date": "2021-04-30",
      "source_url": "https://mhealth.jmir.org/2021/4/e22487/",
      "source_location": "main Results: Accuracy of the Scales",
      "limitation": "Signed scale-minus-reference differences; IQR is not full range or confidence interval; main text chosen over conflicting abstract; different patients per scale; not longitudinal validation",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "synthetic",
      "entity": "invented fat-mass series",
      "visit": 1,
      "reference": 20,
      "device_A": 23,
      "device_B": 20,
      "unit": "kg",
      "limitation": "Not observed participants or real devices; A has constant +3 kg error, B misses change",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "synthetic",
      "entity": "invented fat-mass series",
      "visit": 2,
      "reference": 19,
      "device_A": 22,
      "device_B": 20,
      "unit": "kg",
      "limitation": "Not observed participants or real devices; A has constant +3 kg error, B misses change",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "synthetic",
      "entity": "invented fat-mass series",
      "visit": 3,
      "reference": 18,
      "device_A": 21,
      "device_B": 20,
      "unit": "kg",
      "limitation": "Not observed participants or real devices; A has constant +3 kg error, B misses change",
      "checked_date": "2026-11-20"
    },
    {
      "record_type": "synthetic_calculation",
      "entity": "visit 3 minus visit 1",
      "reference": -2,
      "device_A": -2,
      "device_B": 0,
      "unit": "kg",
      "formula": "last-first",
      "limitation": "Only explains why agreement and responsiveness differ",
      "checked_date": "2026-11-20"
    }
  ]
}